HET
OPERATIONS ENGINEER MANUAL: SAMS STARTUP
AND CALSCREEN CHECK (REV08.08.04)
General
Description:
Startup the SAMS software
1.0 Start SAMS Server.
1.1 Place the cursor in the left
most monitor, second screen,
then press and hold down the right mouse button.
1.2 A menu will appear. Select ‘SAMS’ from the menu; then from the
submenu select ‘Start SAMS Server - Gap’ and release the right mouse
button. A shell will appear.
1.3 Labview will start, and in turn,
automatically start and run SAMS Server .
1.4 Watch the status bar at the
bottom of the SAMS Server interface until the message “Running Main Application
Loop” appears.
1.5 Go back to the Jove menu and
select ‘Start SAMS GUI’ The SAMS GUI
interface will appear.
1.6 From SAMs
GUI select “Operate” from Change Mode This concludes the SAMS
software startup.
Fault
Conditions and Solutions:
metric is steadily
growing
PMCS
new combo and both PMCS and TSPS busy lights show constant orange
SPS out of memory (see SPS section)
SAMS/PMC
loop not cycling
check .ttp values in SAMS
if sams sending all zeros & 7 sec from get ttp to ttp done
the needed
correction size may exceed the max setting
try opening the
max move size a bit
if sending all
zeros, & 1 sec ... ? dunno
metric exceedingly
large
check edge sensor
data
if entire bank
(see mirror diagram) gives values of -384.00
open SPS/SAMS loop
(select all no's in PMUI - SAMS Gui)
quit/close all
SAMS software
power cycle brick
(in UER)
check MTM master
timing module in igloo (goes to brick in UER)
3 channels - all hould
be pulsed, not steady
if problem
persists a hub board may be bad - SAMS deadlined
bring SAMS
software back up
reclose
SPS/SAMS loop
sensor reported bad
open loop
return to standby
remove in SAMS
back to operate
wait a few minutes
check for error msg's
if none
reclose
loop
SAMS
???
server halted (when starting labview)
~10
minutes did not elapse since old version of SAMS was closed.
wait and retry.