Changes between Version 136 and Version 137 of HetProcedures/RA/hpf
- Timestamp:
- Oct 21, 2018 10:19:09 PM (6 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
HetProcedures/RA/hpf
v136 v137 115 115 {{{ hpframp AD_Leo 30 2 }}} 116 116 117 You can set the type to Eng so that they are not science exposures and sent to the TACC by doing a {{{ tims detect pyhxrg:SetObstype:Eng }}} 118 117 119 === Science observing === #obs 118 120 … … 163 165 164 166 {{{ restart_detectorclient }}} is called daily as part of the crontab at 16:45 UT to restart the HPF client daily. It can 165 also be called on demand (e.g. to deal with one of the FIFO Errorfailure modes that crop up occasionally).167 also be called on demand. 166 168 167 169 There are two new executable scripts: … … 240 242 ...lakeshore336 241 243 ......Success 242 Sending Cli16420: Detect Cli1642 0 message:Detect_shuttingdown244 Sending Cli16420: Detect Cli1642 * tims detect pyhxrg:SetObstype:Sci0 message:Detect_shuttingdown 243 245 Shutting down... 244 246 Connection lost to: lakeshore336 … … 274 276 == HPF twilight sky flats (evening) == 275 277 276 ( These are not regularly observed. Good flats were taken in Oct 2018.)278 (Not required but desired when it does not interfere with other operations). 277 279 278 280 Open the PFIP shutter. … … 282 284 Starting ~3min after sunset: 283 285 286 * tims detect pyhxrg:SetObstype:Cal 284 287 * open PFIP shutter 285 288 * hpframp twilight 28 1 286 289 * close PFIP shutter 290 * tims detect pyhxrg:SetObstype:Sci 287 291 288 292 After that finishes, close the PFIP shutter before attempting LRS2 flats.