    479479=== Fixing the calibration connection ===
    480480
    481 tims Calib shutdown
    482 
    483 cntrl-a n to move to next times window
    484 
    485 python -m TIMS.clients.tims_calib
    486 
    487 
    488 
    489 
     481If the etalon turn off command hangs or any of the cal lamps try the following:
     482
     483{{{ tims Calib shutdown }}} in any hpf window with a prompt
     484
     485In the TIMS HET window do a '''cntrl-a n''' five times to move to screen 2 which is the TIMS Calib screen.   It should have a prompt.  If it does not you may need to do a few cntrl-c to kill any hanging processes.
     486
     487In the TIMS Calib screen you should do an up arrow and see the {{{ python -m TIMS.clients.tims_calib }}} command which restarted the Calib client.   Wait a few minutes for it to finish.   
     488
     489Take a quick lfc image and perhaps a quick sky exposure to make sure all of the shutters and baffles are working.
     490