    11= Troubleshooting =
    22
     3== Missing mirrors in the reference measurement of AOA ==
     4
     5The ultimate goal here is for the mirror locations for each mirror to be
     6in a file called mirror_locations.txt on the wave computer in the location
     7C:\aoa\mirror_locations.txt.
     8
     9Step 1:  Obtain a spot image of the reference grid.
     10- under the wavescope software go to the menu selection
     11  Alignment->Manual, select the "Show Spot Image" under the
     12  Reference Source.
     13- snap an image using the command "snap <filename>"
     14  (this will save the image in E:<filename>.fits)
     15
     16Step 3:  Check spot locations
     17- ds9 /home/mars/guider/wave/<filename>.fits
     18- see if the spot locations are close to existing mirror_locations.txt.
     19- you may want to copy over the mirror_locations.txt to E as well to make this go faster.
     20
     21Step 4: Create new spot locations if needed
     22- (iraf stuff)
     23
     24Step 5: Copy the new mirror_locations.txt to the AOA software directory
     25- place a copy of the new file in /home/mars/guider/wave/mirror_locations.txt and /home/mars/guider/wave/mirror_locations_YYYYMMDD.txt
     26- go to the wave computer, open up the shortcut to aoa