Troubleshooting

Missing mirrors in the reference measurement of AOA

The ultimate goal here is for the mirror locations for each mirror to be in a file called mirror_locations.txt on the wave computer in the location C:\aoa\mirror_locations.txt.

Step 1: Obtain a spot image of the reference grid.

  • under the wavescope software go to the menu selection Alignment->Manual, select the "Show Spot Image" under the Reference Source.
  • snap an image using the command "snap <filename>" (this will save the image in E:<filename>.fits)

Step 3: Check spot locations

  • ds9 /home/mars/guider/wave/<filename>.fits
  • see if the spot locations are close to existing mirror_locations.txt.
  • you may want to copy over the mirror_locations.txt to E as well to make this go faster.

Step 4: Create new spot locations if needed

  • (iraf stuff)

Step 5: Copy the new mirror_locations.txt to the AOA software directory

  • place a copy of the new file in /home/mars/guider/wave/mirror_locations.txt and /home/mars/guider/wave/mirror_locations_YYYYMMDD.txt
  • go to the wave computer, open up the shortcut to aoa
  • rename the current mirror_locations.txt to mirror_locations.txt-old
  • navigate to the E: drive, copy the two new files to the aoa directory

Step 6: restart the wavescope software if needed.